ANALYTICAL E MODULE

Received Dec 16, 2009 · Event occurred Dec 2, 2000

Report 1823260-2009-08504 · MDR key 1564501

Device

Generic name

Immunochemistry Analyzer- Jje

Manufacturer

Roche Diagnostics

Model number

E MODULE

Catalog number

03739040001

Product problems

  • High Test Results

Patient

33 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

CUSTOMER REPORTED AN ERRONEOUS FOLATE RESULT OF 19.01 NG PER ML FOR ONE PATIENT SAMPLE. SAME SAMPLE WAS RUN ON A COBAS E601 ANALYZER WHICH RECOVERED 6 NG PER ML. ON (B) (6) 2009, CUSTOMER RAN SAME SAMPLE ON AN E170 ANALYZER WHICH RECOVERED 5.97 NG PER ML. THE SAMPLE TYPE WAS SERUM. PATIENT WAS NOT TREATED OR ADVERSELY AFFECTED DUE TO THE ORIGINAL RESULT. THE FIELD SERVICE REPRESENTATIVE WAS UNABLE TO DETERMINE A CAUSE. TO VERIFY ANALYZER PERFORMANCE, HE COMPLETED PERFORMANCE TESTS WITH NO DISCREPANCIES.