ELECSYS CMV IGM
Received Oct 19, 2022 · Event occurred Sep 20, 2022
Report 1823260-2022-03250 · MDR key 15632292
Device
Generic name
Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus
Manufacturer
Roche DiagnosticsModel number
NACatalog number
CMV IGM ELECSYS E2G
Lot number
63027501
Product problems
- False Negative Result
- False Negative Result
Patient
25 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE FALSE NEGATIVE CMV IGM ELECSYS RESULTS FROM COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER (B)(4). FOR A SAMPLE COLLECTED ON (B)(6) 2022, THE RESULT WAS NEGATIVE. WHEN TESTED ON A VIDAS ANALYZER, THE RESULT WAS POSITIVE. FOR A SAMPLE COLLECTED ON (B)(6) 2022, THE RESULT WAS NEGATIVE. WHEN TESTED ON A VIDAS ANALYZER, THE RESULT WAS POSITIVE. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Additional Manufacturer Narrative
THE SAMPLES IN QUESTION WERE SUBMITTED FOR INVESTIGATION. FOR THE (B)(6) 2022 SAMPLE: THE CMV IGM RESULT WAS 0.466 COI (NON-REACTIVE). THE CMV IGG RESULT AFTER DILUTION WAS 1724 U/ML (REACTIVE). THE CMV IGG AVIDITY WAS 60.9%. FOR THE (B)(6) 2022 SAMPLE: THE CMV IGM RESULT WAS 0.377 COI (NON-REACTIVE). THE CMV IGG RESULT AFTER DILUTION WAS 1612 U/ML (REACTIVE). THE CMV IGG AVIDITY WAS 61.6%. THE CUSTOMER'S RESULTS WERE REPRODUCIBLE. A GENERAL REAGENT ISSUE CAN BE EXCLUDED. THE INVESTIGATION IS ONGOING.
Additional Manufacturer Narrative
THE CALIBRATION AND QC DATA WERE WITHIN THE EXPECTED RANGES. ADDITIONAL ANALYSIS SHOWED THE SAMPLES CONTAINED CMV-SPECIFIC IGG ANTIBODIES OF HIGH AVIDITY, SUGGESTING THAT OTHER METHODS MIGHT DETECT PERSISTING CMV IGM ANTIBODIES IN THE SAMPLES. THE OBTAINED RESULTS ARE NOT INDICATIVE OF AN EARLY PHASE OF INFECTION WITH CMV. MEDWATCH FIELDS A2 - AGE AT TIME OF EVENT, A3 - SEX, AND B7 - OTHER RELEVANT HISTORY WERE UPDATED.