inforMED
MalfunctionKPN

CENTRELLA WITH WATCHCARE

Received Oct 11, 2022 · Event occurred Jul 1, 2022

Report MW5112537 · MDR key 15588249

Device

Generic name

Alarm, Conditioned Response Enuresis

Manufacturer

Hill-rom, Inc.

Catalog number

B7900B

Product problems

  • Device-Device Incompatibility

Patient

Prefer Not To Disclose

  • Non specific EKG/ECG Changes
  • Non specific EKG/ECG Changes

Narrative

Description of Event or Problem

INTERFERENCE FROM NEW HILL ROM BED WATCHCARE INCONTINENCE MANAGEMENT SYSTEM CAUSING SPIKES IN PATIENT ECG IN NEW GE CARESCAPE TELEMETRY SYSTEM. HILLROM DEACTIVATED WATCHCARE SYSTEM IN ALL NEW BEDS PROBLEM NOT CORRECTED YET. FDA SAFETY REPORT ID# (B)(4).