inforMED
MalfunctionLOP

COBAS INTEGRA 400

Received Nov 18, 2009 · Event occurred Oct 8, 2009

Report 1823260-2009-07905 · MDR key 1553454

Device

Generic name

Clinical Chemistry Analyzer - Jje

Manufacturer

Roche Diagnostics

Model number

I400

Catalog number

21045199001

Product problems

  • High Test Results

Patient

UNK

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

THE USER STATED THEY HAVE HAD "AT LEAST 5 OCCURRENCES" OF A HIGH HEMOGLOBIN A1C RESULT WHICH WAS REPORTED, THEN REPEATED AS NORMAL. THE DOCTORS HAVE STATED THE PATIENTS WITH THE HIGH RESULTS HAVE NO OTHER INDICATION THAT WOULD CAUSE A HIGH RESULT. THE PATIENTS HAD NOT BEEN TREATED AS "THE DOCTORS FELT THE PATIENTS HAD NOT BEEN THAT HIGH AND NOTIFIED THE LAB, RESULTING IN THE REPEAT TESTING." THE USER PROVIDED THREE EXAMPLES. ON (B)(6)2009, A SAMPLE FOR A FEMALE PATIENT RESULTED 13.6% INITIALLY AND 7.56% UPON REPEAT. ON (B)(6)2009, A SAMPLE FOR A FEMALE PT RESULTED 11.8% INITIALLY AND 6.2% UPON REPEAT. ON (B)(6)2009, A SAMPLE FOR A MALE PATIENT RESULTED 14.1% INITIALLY AND 7.4% UPON REPEAT. THE FIELD SERVICE REPRESENTATIVE COULD NOT FIND A CAUSE AND CHECKED THE ANALYZER BY PERFORMING PIPETTING AND PHOTOMETER CHECKS WITH ACCEPTABLE RESULTS. HE ALSO CHECKED THE CONFIGURATION OF THE ASSAY AND RAN EIGHT SPECIMENS AND COULD NOT DUPLICATE THE ISSUE.