inforMED
InjuryMBB

HIGH VISCOSITY BONE CEMENT

Received Sep 15, 2022 · Event occurred Mar 10, 2015

Report MW5112078 · MDR key 15438086

Device

Generic name

Bone Cement, Antibiotic

Catalog number

6195-1-001

Lot number

424BA86DV

Product problems

  • Loss of or Failure to Bond
  • Product Quality Problem

Patient

68 YR · Female

  • Failure of Implant
  • Failure of Implant

Narrative

Description of Event or Problem

THIS PATIENT HAD A TOTAL KNEE REPLACEMENT PERFORMED ON THE DATE INDICATED. THE KNEE FAILED MUCH EARLIER THAN EXPECTED. WORK UP WAS NEGATIVE FOR INFECTION. THE RECORDS INDICATE THE ORIGINAL SURGERY WAS DONE USING STRYKER HIGH VISCOSITY BONE CEMENT FOR THE KNEE REPLACEMENT SURGERY. THE PATIENT HAD TO HAVE REVISION SURGERY AS INDICATED. THE FINDINGS AT TIME OF SURGERY WERE CLASSIC FOR A CEMENT FAILURE IN THAT THE CEMENT WAS NOT BONDED OR ADHERENT TO THE IMPLANTS BUT WAS AFFIXED TO THE BONE. THIS CASE IS ONE OF MANY CASES OF FAILURE AFTER 1.5-2.5 YEARS WHEN USING THIS PRODUCT. FDA SAFETY REPORT ID# (B)(4).