inforMED
MalfunctionFPL

TI100 TRANSPORT INCUBATOR

Received Feb 27, 1998

Report 2510954-1998-00002 · MDR key 153345

Device

Generic name

Transport Incubator

Model number

TI100

Catalog number

6730070

Lot number

*

Product problems

  • Fire

Patient

NA

Narrative

Description of Event or Problem

IT WAS REPORTED THAT A MODEL TI100 TRANSPORT INCUBATOR WAS IN A HALLWAY PLUGGED INTO AN AC RECEPTACLE TO MAINTAIN CONSTANT WARMTH AND BATTERY CHARGE. THE UNIT CAUGHT FIRE AND WAS SUBSEQUENTLY EXTINGUISHED BY THE FIRE DEPARTMENT.

Additional Manufacturer Narrative

A MFR ENGINEER PERFORMED AN EVAL OF THE DEVICE. ENGINEER'S FINDINGS CONFIRMED THAT THE BATTERY CHARGER CONNECTOR 4P1/4J1 HAD BECOME OVERHEATED, WHICH IN TURN CAUSED OVERHEATING OF ADJACENT COMPONENTS. A MEDICAL DEVICE SAFETY ALERT WAS ISSUED IN 1995 NOTIFYING USERS OF THIS SITUATION. FACILITY INDICATED BEING AWARE OF THIS NOTIFICATION.