MalfunctionCDD

ELECSYS VITAMIN B12 II

Received Aug 31, 2022 · Event occurred Aug 8, 2022

Report 1823260-2022-02655 · MDR key 15328777

Device

Generic name

Radioassay, Vitamin B12

Manufacturer

Roche Diagnostics

Model number

B12 G2

Catalog number

07212771190

Lot number

58858304

Product problems

  • Crack
  • Unsealed Device Packaging
  • Low Test Results
  • Low Readings
  • Crack
  • Unsealed Device Packaging
  • Low Test Results
  • Low Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE CUSTOMER REPLACED THE REAGENT KIT AND THE PATIENT AND QC RESULTS WERE NORMAL. THE CUSTOMER CHECKED THE REAGENT KIT IN QUESTION, AND "A LOT" OF CRYSTALS WERE AT THE MOUTH OF THE BOTTLE OF THE PRE-TREATMENT SOLUTION. THERE WAS NOT A LOT OF LIQUID IN THE BOTTLE SUGGESTING A LEAK. THE REAGENT KIT WAS REQUESTED FOR INVESTIGATION.

Description of Event or Problem

THE INITIAL REPORTER QUESTIONED RESULTS FOR QC AND MULTIPLE PATIENT SAMPLES TESTED WITH ONE PARTICULAR REAGENT KIT OF ELECSYS VITAMIN B12 II (VITAMIN B12 II) ON A COBAS 8000 CORE UNIT. THE CUSTOMER PROVIDED 2 SETS OF DISCREPANT RESULTS. THE FIRST SET: THE INITIAL RESULT WAS 160.9 PMOL/L. THE REPEAT RESULT WAS > 1476 PMOL/L. THE 2ND SET: THE INITIAL RESULT WAS 36.9 PMOL/L. THE REPEAT RESULT WAS 161.1 PMOL/L. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE 8000 CORE UNIT SERIAL NUMBER WAS (B)(4).

Additional Manufacturer Narrative

NO CALIBRATION DATA WAS PROVIDED FOR THE AFFECTED REAGENT KIT. CALIBRATION SIGNALS FROM A DIFFERENT REAGENT KIT WITH THE SAME LOT NUMBER WERE ACCEPTABLE. ON 06-AUG-2022 QC WAS LOW, AND OUT OF RANGE FOR THE AFFECTED REAGENT KIT. AROUND THE SAME TIME, THERE WERE OTHER LOW QC RESULTS. THE REST OF THE QC RESULTS WERE WITHIN THE ACCEPTABLE RANGE. THE REAGENT KIT WAS RECEIVED FOR INVESTIGATION WHERE CALIBRATION COULD NOT BE GENERATED. AN ALARM WAS RECEIVED: "REAGENT PROBE DID NOT DETECT LIQUID LEVEL OF PRETREATMENT." BASED ON A REVIEW OF WORLDWIDE DATA OF QC RECOVERY FOR THE REAGENT LOT USED BY THE CUSTOMER, NO REAGENT ISSUE WAS FOUND. THE INVESTIGATION DETERMINED THE CRACK IN THE NECK OF PRE-TREATMENT SOLUTION BOTTLE 2 CAUSED THE LIQUID LEAKAGE. A GENERAL PRODUCT PROBLEM HAS BEEN EXCLUDED. ONLY SINGLE REAGENT KITS ARE AFFECTED. OTHER REAGENT KITS FROM THE SAME LOT PERFORMED WITHIN SPECIFICATION AT THIS CUSTOMER SITE AND OTHER CUSTOMER SITES. THE SPECIFIC CAUSAL LINK BETWEEN THE BOTTLE DAMAGE AND THE ISSUE WITH THE SINGLE REAGENT KIT COULD NOT BE DETERMINED.

Additional Manufacturer Narrative

THE REAGENT KIT WAS RECEIVED FOR INVESTIGATION. SECTIONS D9 AND H3 WERE UPDATED. THE REAGENT PACKAGING WAS VISUALLY INSPECTED AND THERE WERE WATER STAINS ON THE BOX. THE REAGENT KIT WAS VISUALLY INSPECTED AND THERE WERE NO TRACES OF LEAKAGE OR DAMAGE. THE REAGENT KIT WAS DISASSEMBLED AND THERE WAS NO DAMAGE OR VISIBLE CRACKS ON THE REAGENT BOTTLES OR PRE-TREATMENT SOLUTION BOTTLE 1. THE MOUTH OF PRE-TREATMENT SOLUTION BOTTLE 2 WAS DAMAGED. THE PRE-TREATMENT SOLUTION BOTTLE 2 WAS EMPTY. THE INVESTIGATION IS ONGOING.