inforMED
MalfunctionIOY

SUPPORT ARM 176

Received Aug 23, 2022 · Event occurred Aug 2, 2022

Report 8010042-2022-01462 · MDR key 15274909

Device

Generic name

Support, Arm

Model number

SUPPORT ARM 176

Catalog number

6405976

Product problems

  • Fracture
  • Fracture

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE SUPPORT ARM HOLDING THE PATIENT CIRCUIT BROKE. PATIENT INVOLVEMENT IS UNKNOWN. MANUFACTURER'S REF #: (B)(4).

Additional Manufacturer Narrative

THE REPORTED ISSUE WITH BROKEN SUPPORT ARM WAS CONFIRMED DURING EVALUATION OF THE PROVIDED PICTURE AND IN PROBLEM DESCRIPTION. ACCORDING TO THE PICTURE, THE SUPPORT ARM BROKE AT THE CLAMP ATTACHMENT. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGNED ACCORDING TO STANDARD. THE BROKEN SUPPORT ARM WAS NOT RETURNED TO US FOR FURTHER INVESTIGATION. THE ROOT CAUSE OF THE BROKEN SUPPORT ARM HAS NOT BEEN CONCLUSIVELY DETERMINED. HOWEVER; THE BRAKING POINT INDICATES THAT IT MOST LIKELY HAS BEEN EXPOSED TO A MECHANICAL FORCE GREATER THAN IT IS DESIGNED TO SUSTAIN.