ARCHITECT CYCLOSPORINE REAGENT KIT
Received Jul 25, 2022 · Event occurred Jul 13, 2022
Report 1415939-2022-00055 · MDR key 15099869
Device
Generic name
Cyclosporine
Manufacturer
Abbott LaboratoriesModel number
1L75-25Catalog number
01L75-25
Lot number
37701FP00
Product problems
- Low Test Results
- Low Test Results
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE CUSTOMER OBSERVED A FALSELY DECREASED ARCHITECT CYCLOSPORINE RESULT FOR ONE PATIENT. THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT, AFTER RECEIVING MEDICATION ON 11JUL, WAS 872.5, REPEAT, ON 13JUL, WAS 4.3 NG/ML. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. PHONE COMPLETE ENTRY(B)(6).
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION FOR A FALSELY DECREASED ARCHITECT CYCLOSPORINE RESULT INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TICKET TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND TESTING OF THE COMPLAINT LOT NUMBER. TRENDING REVIEW DETERMINED NO RELATED TREND FOR THE PRODUCT. RETURN TESTING WAS NOT COMPLETED, AS RETURNS WERE NOT AVAILABLE. ACCURACY TESTING WAS PERFORMED USING AN IN-HOUSE RETAINED KIT OF LOT 37701FP00, STORED AT THE RECOMMENDED STORAGE CONDITION. ALL SPECIFICATIONS WERE MET INDICATING THAT THE LOT IS PERFORMING ACCEPTABLY. MANUFACTURING DOCUMENTATION FOR THE LIKELY CAUSE LOT WAS REVIEWED AND DID NOT IDENTIFY ANY ISSUES. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH THE LIKELY CAUSE LOT AND COMPLAINT ISSUE. LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER'S ISSUE. BASED ON THE INFORMATION PROVIDED AND ABBOTT DIAGNOSTICS¿ COMPLAINT INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF ARCHITECT CYCLOSPORINE REAGENT, LOT NUMBER 37701FP00, WAS IDENTIFIED.