OR1 NEO IP, ENH
Received Jul 22, 2022 · Event occurred Jun 24, 2022
Report 9610617-2022-00129 · MDR key 15085241
Device
Generic name
Neo Ip Controller
Manufacturer
Karl Storz Se & Co. KgModel number
WUIS2546-01Catalog number
WUIS2546-01
Product problems
- No Display/Image
- No Display/Image
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
OUR ONSITE TECHNICIAN WAS UNABLE TO REPRODUCE THE ISSUE, AND REPORTED THAT THE SITE DID HAVE A GROUNDING ISSUE ON A 120V RECEPTACLE IN THIS ROOM THAT HAS SINCE BEEN REPAIRED AFTER INCIDENT. CAUSE IS UNKNOWN AS WE WERE UNABLE TO REPLICATE THE ISSUE. SINCE THE ESU IS A ROLLING UNIT, THE SITE HAS BEEN INSTRUCTED TO ISOLATE THE MACHINE TO INCLUDE ALL CABLES AND UNIT FOR FURTHER EXTENSIVE TESTING IF THE PROBLEM OCCURS AGAIN.
Description of Event or Problem
CUSTOMER REPORTED THAT WHEN TRIGGERING THE ESU-ERBE VIO 300D, THE OLYMPUS CV-190 CAMERA IMAGE WOULD GO BLACK WHEN ROUTED TO ANY MONITOR/DISPLAY, INCLUDING VITALS. THE PATIENT NEEDED TO BE MOVED TO A DIFFERENT ROOM AND SURGEON WAS ABLE TO COMPLETE THE PROCEDURE; THERE WAS NO NEGATIVE PATIENT IMPACT.