LUCEA 40
Received Jul 22, 2022 · Event occurred May 23, 2022
Report 9710055-2022-00291 · MDR key 15081665
Device
Generic name
Device, Medical Examination, Ac Powered
Manufacturer
Maquet SasModel number
ARDLCA109001ACatalog number
ARD568601997
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. DEVICE NOT RETURNED TO MANUFACTURER.
Description of Event or Problem
ON (B)(6) 2022 GETINGE BECAME AWARE OF AN ISSUE WITH ONE OF SURGICAL LIGHTS - LUCEA 40. IT WAS STATED THE COVER OF THE HEADLIGHT CRACKED AND THE PARTICLES WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY DETACHMENT OF PART OR PARTICLES MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT.
Additional Manufacturer Narrative
ACCORDING TO THE REPORTING TIMEFRAME WE WOULD LIKE TO PROVIDE THE INFORMATION ABOUT CURRENT STATUS OF THE ISSUE. PLEASE BE ADVISED THAT IT IS BEING INVESTIGATED. ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. THE CORRECTION OF CATALOG# FIELD DEEM REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. #D4: PREVIOUS CATALOG #: ARDLCA109001A. CORRECTED CATALOG#: ARD568601997.
Description of Event or Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).
Additional Manufacturer Narrative
ACCORDING TO THE REPORTING TIMEFRAME WE WOULD LIKE TO PROVIDE THE INFORMATION ABOUT CURRENT STATUS OF THE ISSUE. PLEASE BE ADVISED THAT IT IS BEING INVESTIGATED. ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION. THE CORRECTION OF EVENT SITE CITY#, EVENT SITE ADDRESS# AND EVENT SITE POSTAL CODE# FIELDS DEEM REQUIRED. THIS IS BASED ON THE INTERNAL EVALUATION. PREVIOUS EVENT SITE CITY: (B)(6). CORRECTED EVENT SITE CITY: (B)(6). PREVIOUS EVENT SITE POSTAL CODE: (B)(6). CORRECTED EVENT SITE POSTAL CODE: (B)(6). PREVIOUS EVENT SITE ADDRESS: (B)(6). CORRECTED EVENT SITE ADDRESS: (B)(6).
Description of Event or Problem
MANUFACTURER'S REFERENCE NUMBER OT (B)(4).
Additional Manufacturer Narrative
GETINGE BECAME AWARE OF AN ISSUE WITH SURGICAL LIGHTS - LUCEA 40. IT WAS STATED THE COVERS OF THE HEADLIGHTS CRACKED AND SOME PARTICLES WERE MISSING. THERE WAS NO INJURY REPORTED, HOWEVER, WE DECIDED TO REPORT THE ISSUE IN ABUNDANCE OF CAUTION AS ANY DETACHMENT OF PART OR PARTICLES MAY LEAD TO POTENTIAL INFECTION OF THE PATIENT. BASED ON THE INFORMATION COLLECTED, IT WAS ESTABLISHED THAT WHEN THE EVENT OCCURRED, SURGICAL LIGHTS DID NOT MEET THEIR SPECIFICATION, SINCE CRACKS IN COVERS, RESULTING IN MISSING PLASTIC PARTICLES OR DETACHMENT OF COVERS COULD BE CONSIDERED AS A TECHNICAL DEFICIENCY, AND IN THIS WAY DEVICES CONTRIBUTED TO THE EVENT. DEVICES WERE BEING USED FOR PATIENT DIAGNOSIS WHEN THE EVENT OCCURRED. WHEN REVIEWING REPORTABLE EVENTS FOR THIS TYPE OF ISSUE WE WERE ABLE TO ESTABLISH THAT THE RECEIVED INCIDENTS ARE OCCURRING AT A MODERATE RATIO FOR THE BROKEN COVERS. WE HAVE BEEN ABLE TO CONFIRM THAT THE INVESTIGATED ISSUE HAS NEVER LED TO SERIOUS INJURY OR WORSE, TO OUR KNOWLEDGE. AS PER EXPERTISE PERFORMED BY THE SUBJECT MATTER EXPERT AT MANUFACTURING SITE THE CRACKS DETECTED ON THE HANDLE INTERFACE AND TRANSPARENT COVER FIXING POINTS WERE PROBABLY CAUSED BY IMPROPER USE, COLLISIONS, INCOMPATIBLE CLEANING PRODUCTS OR PROTOCOL. TO PREVENT ANY INCIDENT THE USER MANUAL FOR LUCEA 10/40 (IFU 01701 EN 12) ON PAGE 22 MENTIONS TO CHECK THE INTEGRITY OF THE LIGHT HEADS DURING DAILY INSPECTION. FURTHERMORE, THE SAME USER MANUAL ON PAGES 29-30 EXPLAINS HOW TO CLEAN AND DISINFECT THE LIGHTHEADS. THIS DOCUMENT INCLUDES SOME RECOMMENDED PRODUCTS AND SOME PROHIBITED PRODUCTS. IN ORDER TO AVOID TORQUES APPLIED ON THE TRANSPARENT HOUSING DURING USE THE USER MANUAL FOR LUCEA 10/40 (IFU 01701 EN 12) ON PAGE 25 MENTIONS TO HANDLE THE LIGHT HEAD BY THE HANDLE. WE BELIEVE THE RELATED DEVICES ARE PERFORMING CORRECTLY IN THE MARKET. WE ALSO BELIEVE THAT IF THE MANUFACTURER¿S RECOMMENDATION HAD BEEN FOLLOWED THE INCIDENT COULD HAVE BEEN AVOIDED. GETINGE SHALL CONTINUE TO MONITOR FOR ANY FURTHER EVENTS OF THIS NATURE AND DO NOT PROPOSE ANY FURTHER ACTION AT THIS TIME.
Description of Event or Problem
MANUFACTURER'S REFERENCE NUMBER (B)(4).