PENTAX
Received Jul 21, 2022 · Event occurred Jun 24, 2022
Report 9610877-2022-55879 · MDR key 15071450
Device
Generic name
Video Sigmoidoscope 4.2c 12.8t 700l Fwj
Manufacturer
Hoya Corporation Pentax Tokyo OfficeModel number
ES-3870KProduct problems
- Optical Distortion
- Optical Distortion
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE OBJECTIVE LENS UNIT CLOUDY (NOT CLEAR VIEW). BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE MOISTURE CONDENSATION IN THE OBJECTIVE LENS UNIT. IN ADDITION, WE CONFIRMED THAT THE INSERTION FLEXIBLE TUBE (IFT) COATING DAMAGE, THE WATER NOZZLE CLOGGED, THE INSERTION FLEXIBLE TUBE (IFT) BUMP, THE ANGLE WIRE GRINDING, AND THE AIR/WATER SOCKET CHIPPED/CUT O-RING; HOWEVER, THEY ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT "HR-RPT-0586(IMAGE FAILURE)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE(CLOUDY).