Received Oct 8, 2009 · Event occurred Sep 8, 2009

Report 1836145-2009-00003 · MDR key 1506623

Device

Generic name

Ceiling Column

Model number

680

Product problems

  • Unintended System Motion

Patient

UNK

  • Head Injury
  • Headache

Narrative

Description of Event or Problem

ALLEGED THE RETRACTABLE CEILING COLUMN DRIFTS DOWN UNEXPECTEDLY. ON (B)(6), AN OR EMPLOYEE WAS HIT ON THE HEAD WHEN THE CEILING COLUMN DRIFTED DOWN. THE EMPLOYEE HAD A HEADACHE, WENT TO ER, IS REPORTED TO HAVE A CONCUSSION AND WAS SENT HOME FOR THE DAY. THE EMPLOYEE IS WORKING BUT STILL CLAIMS TO HAVE AN ONGOING HEADACHE. THE FACILITY INDICATED THE DRIFT HAS BEEN HAPPENING FOR SOMETIME.

Additional Manufacturer Narrative

THE TECHNICIAN INSPECTED THE COLUMN AND COULD NOT DUPLICATE THE ALLEGED MALFUNCTION. THE TECHNICIAN DID FIND SMALL AIR LEAKS BETWEEN THE CONTROL HOSES AND PRESS LOCK FITTINGS. THE HOSES WERE RE-FITTED TO THE PRESS LOCK FITTINGS TO STOP THE LEAKS. THE ACCOUNT DECLINED TO PROVIDE ANY ADDITIONAL INFORMATION REGARDING THE EVENT.