inforMED
MalfunctionDWH

SAGITTAL BLADE

Received Sep 17, 2009 · Event occurred Aug 17, 2009

Report 9616696-2009-00137 · MDR key 1505026

Device

Generic name

Saw Blade & Accessories

Catalog number

2108302000

Lot number

UNK

Product problems

  • Break
  • Detachment Of Device Component

Patient

UNK

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE BLADE PRODUCT SUBJECT TO THIS MDR, BROKE AT THE MOUNT TAB DURING A TOTAL KNEE PROCEDURE. IT WAS FURTHER REPORTED THAT THE BLADE CAME OUT OF THE HANDPIECE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS ALLEGED ISSUE.

Additional Manufacturer Narrative

LOT NUMBER INFO HAS BEEN SUPPLIED SO AS TO PERMIT FURTHER INVESTIGATION OF THIS POTENTIAL ISSUE. PRODUCT ALLEGEDLY INVOLVED IN THIS ISSUE WAS VISUALLY INSPECTED. IT IS CONFIRMED THAT ONE TAB HAS BROKEN OFF OF THE MOUNT OF THE BLADE. SCORE MARKINGS ON THE BLADE INDICATE THAT IT WAS INSERTED CORRECTLY IN TO THE HANDPIECE USED.