inforMED
MalfunctionGDZ

360 DEGREE SCALPEL HANDLE

Received Jul 13, 2022 · Event occurred Jun 9, 2022

Report 1416605-2022-00008 · MDR key 15008798

Device

Generic name

Scalpel Handle

Model number

K360

Catalog number

K360

Lot number

0322

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

Unknown

  • Laceration(s)
  • Laceration(s)

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE WHICH INCLUDED USE OF A 360 DEGREE SCALPEL HANDLE, PART OF THE HANDLE DETACHED AND THE PATIENT OBTAINED A CUT.

Additional Manufacturer Narrative

HU-FRIEDY HAS CONTACTED THE CUSTOMER TO OBTAIN ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT AND REQUESTED THE DEVICE TO BE RETURNED FOR EVALUATION. THE INVESTIGATION INTO THE REPORTED EVENT IS IN PROCESS; A FOLLOW-UP MDR WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

BASED ON THE INVESTIGATION PERFORMED BY HU-FRIEDY, THE TIP OF THE 360 DEGREE SCALPEL HANDLE SEPARATED DUE TO PRODUCTION PERSONNEL OVER BUFFING THE HANDLE DURING THE PRESS FITTING PROCESS. THE DEVICE SUBJECT OF THE EVENT WAS RETURNED TO HU-FRIEDY.