inforMED
MalfunctionJOS

LIGHTWAVE SUCTION ABLATOR

Received Sep 4, 2009

Report 1017294-2009-00128 · MDR key 1497409

Device

Generic name

Electrode, Electrosurgical

Manufacturer

Conmed Linvatec

Model number

NA

Catalog number

IA-2000-S

Lot number

0901262

Product problems

  • Self-Activation or Keying

Patient

NK

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

THE CUSTOMER REPORTED THAT THIS ABLATOR STARTED TO OPERATE ON ITS OWN WITHOUT DEPRESSING THE BUTTONS. THERE WAS NO INJURY AND MINIMAL SURGICAL DELAY AS A RESULT OF THIS EVENT. THE PROCEDURE WAS COMPLETED AS INTENDED.

Additional Manufacturer Narrative

INVESTIGATION FINDING: CONMED LINVATEC RECEIVED THIS DEVICE FOR EVAL. AN INVESTIGATION FOUND THE ABLATE BUTTON FUNCTIONAL AND THE COAG BUTTON INTERMITTENT. UPON DISASSEMBLY, THE PC BOARD WAS FOUND DISCOLORED WITH EVIDENCE OF CORROSIVE LIKE DEPOSITS ON THE COAGULATION CONTACT AND DOME SWITCH. IN ADDITION, MINIMAL ADHESIVE WAS FOUND ON THE HANDLE HOUSING AT THE BOND SEAM. THIS FAILURE MODE REMAINS UNDER INVESTIGATION.