inforMED
MalfunctionIPF

7550

Received Aug 27, 2009 · Event occurred Jul 28, 2009

Report 1022819-2009-00176 · MDR key 1495627

Device

Generic name

Ipf, Gzj, Gzi, Img

Manufacturer

Chattanooga Group

Model number

7550

Catalog number

7550

Product problems

  • Power Conditioning Problem

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

NON INJURY INCIDENT. COMPLAINANT STATES THAT STIMULATION WILL JOLT WHENEVER PRE-MODIFIED WAVEFORM IS SELECTED.

Additional Manufacturer Narrative

CHATTANOOGA GROUP ENGINEERING EVAL OF THE DEVICE FINDS NO ANOMALIES ON CHANNEL 1 & 2 OUTPUT. CHANNEL 3 & 4 WAS FOUND TO HAVE SHORTED CIRCUIT. DEVICE WAS REPAIRED AND RE-TESTED WITH NO FURTHER ANOMALIES FOUND. CHATTANOOGA GROUP IS PERFORMING FURTHER ANALYSIS TO DETERMINE IF THE SEVERITY AND FREQUENCY OF THIS OCCURRENCE IS WITHIN EXPECTED LEVELS.