Device

Generic name

Body Plethysmograph With Diffusing Capacity

Manufacturer

Medgraphics

Model number

1085-D

Catalog number

NI

Lot number

NI

Product problems

  • Measurements, inaccurate
  • Blood gas measurements, erroneous
  • Low Readings

Patient

NA

Narrative

Description of Event or Problem

IN 2/97, THREE MEDGRAPHICS 1085 ADVANTAGE PLUS SYSTEMS WERE PURCHASED BY MEDICAL CENTER TO REPLACE THE OLDER 1070 MODELS. IN 6/97, IT WAS NOTED THAT THE 1085 WAS PRODUCING COMPARATIVE DIFFUSION READINGS 5-10% LOWER THAN THE 1070. IN 11/97, G.C. COLUMNS WERE REPLACED IN THE THREE 1085 UNITS. THE 1085 UNITS CONTINUED TO PRODUCE RESULTS THAT WERE LOWER THAN THE 1070 BEYOND THE ACCEPTABLE VARIATION (6%).