inforMED
MalfunctionIPF

LEGEND XT 4CH STIM PKG US STD

Received Jun 27, 2022 · Event occurred Jun 1, 2022

Report 9616086-2022-00016 · MDR key 14837167

Device

Generic name

Stimulator, Muscle, Powered

Manufacturer

Djo Llc

Model number

2786

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

Unknown

  • Superficial (First Degree) Burn
  • Superficial (First Degree) Burn

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT THE THE UNIT ALLEGEDLY CAUSED A 1ST DEGREE BURN TO PATIENT DURING STIM TREATMENT. AFTER PERFORMING FUNCTIONAL AND ELECTRICAL TESTING, THE TECHNICIAN WAS UNABLE TO DUPLICATE THE REPORTED COMPLAINT. THE PRODUCT MET MANUFACTURER SPECIFICATIONS. POSSIBLE ROOT CAUSES INCLUDE MOISTURE OR MISAPPLICATION OF THE ELECTRODES.

Description of Event or Problem

IT WAS REPORTED THAT THE THE UNIT ALLEGEDLY CAUSED A BURN TO PATIENT DURING STIM TREATMENT.