inforMED
InjuryLMD

OR1 NEO IP, ENH2

Received Jun 17, 2022 · Event occurred May 24, 2022

Report 9610617-2022-00097 · MDR key 14733266

Device

Generic name

Neo Ip Controller

Model number

WUIS2548

Catalog number

WUIS2548

Product problems

  • Temperature Problem
  • Temperature Problem

Patient

Male

  • Partial thickness (Second Degree) Burn
  • Partial thickness (Second Degree) Burn

Narrative

Description of Event or Problem

CUSTOMER REPORTED A UPS OVERHEATING ALARM ON THE INTEGRATION RACK, AND WHEN THE BIOMED WAS INSPECTING THE EQUIPMENT, HE REACHED ACROSS THE BACK OF THE EQUIPMENT ON THE RACK AND RECEIVED A 2ND DEGREE BURN ON HIS ARM.

Additional Manufacturer Narrative

THE EQUIPMENT WILL BE INSPECTED ONSITE, AND A SUPPLEMENT REPORT WILL BE SUBMITTED WHEN FURTHER INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative

OUR TECHNICIAN WAS UNABLE TO INSPECT THE EQUIPMENT, AS THE CUSTOMER DECIDED THAT THEY DID NOT WANT AN ONSITE VISIT FOR THIS CASE. THEREFORE, A CAUSE FOR THE OVERHEATING ISSUE COULD NOT BE DETERMINED.