MalfunctionJIL

CHEMSTRIP 7

Received Jun 13, 2022 · Event occurred Apr 27, 2022

Report 1823260-2022-01716 · MDR key 14676398

Device

Generic name

Urine Test Strip

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

11008552160

Lot number

53644502

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • High Readings
  • Incorrect, Inadequate or Imprecise Result or Readings
  • High Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THIS REPORT WAS INITIALLY PROVIDED TO ROCHE BY FDA VIA MEDWATCH (B)(4). ROCHE URINE DIPSTICK PRODUCTS ARE GIVING TRACE PROTEIN RESULTS ON A NORMAL QUALITY CONTROL. (EXPECTED RESULT IS NEGATIVE). CHEMSTRIP 7 (PRODUCT NUMBER (B)(4) ACROSS MULTIPLE LOT NUMBERS, LOCATIONS AND USERS. ALSO OBSERVED IN THE CHEMSTRIP 10, AND CHEMSTRIP 10SG PRODUCT. FDA SAFETY REPORT ID #:(B)(4).

Additional Manufacturer Narrative

THE CUSTOMER DROPPED THE QC SOLUTION ONTO THE TEST PADS, TIPPED THE STRIP ALLOWING THE EXCESS SOLUTION TO RUN OFF THE SIDES, AND THEN READ IT. THE RECOMMENDED TECHNIQUE IS TO DIP THE STRIPS INTO THE QC SOLUTION. THE RETENTION MATERIAL OF LOT 53644500, 55018600, AND 51951500 WERE INVESTIGATED AND MEASURED BY VISUAL READING. NO FALSE POSITIVE RESULTS WERE OBSERVED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.