Received May 19, 2009 · Event occurred Apr 15, 2009

Report 3006219967-2009-00002 · MDR key 1461109

Device

Generic name

Endoworks

Model number

IBM 6218-1AU

Catalog number

HW02-102

Product problems

  • Device Emits Odor
  • Loss of Power

Patient

Not reported

  • No Information

Narrative

Description of Event or Problem

IT WAS REPORTED BY THE CUSTOMER (BIOMED) THAT, DURING A COLONOSCOPY EXAM, A BURNING SMELL WAS NOTICED EMANATING FROM THE ENDOWORKS (IBM) COMPUTER. THE COMPUTER LOST POWER AND FURTHER IMAGES COULD NOT BE CAPTURED. THE PROCEDURE WAS COMPLETED WITH THE REMAINING OLYMPUS ENDOSCOPY EQUIPMENT WITHOUT FURTHER INCIDENT.

Additional Manufacturer Narrative

THE HOSPITAL BIOMED EVALUATED THE COMPUTER AFTER THE EXAM AND REPORTED THAT THE POWER SUPPLY HAD FAILED. THE COMPUTER WAS REPAIRED BY THE HOSP BIOMED. THE COMPUTER IS NOT BEING RETURNED BY THE HOSP FOR FURTHER EVAL.

Remedial action

  • Repair