DUAL LUMEN 3.5FR POLYURETHANE UMBILICAL CATHETER
Received Jun 2, 2022 · Event occurred Apr 20, 2022
Report 3007697249-2022-00001 · MDR key 14576388
Device
Generic name
Umbilical Catheter
Manufacturer
Footprint Medical Inc.Model number
P2UVC3.5Lot number
220125
Product problems
- Break
- Stretched
- Break
- Stretched
Patient
5 DA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE UMBILICAL CATHETER WAS REPORTED TO HAVE BEEN PLACED IN THE UMBILICAL VEIN OF A NEWBORN BABY. AFTER SUCCESSFULLY COMPLETING THE COURSE OF THERAPY, THE CATHETER WAS REMOVED. UPON EXTRACTION, THE NNP NOTICED THAT THE TIP APPEARED JAGGED AND STRETCHED. CONCERNED THAT THE TIP MAY HAVE BROKEN OFF IN THE PATIENT, AN X-RAY WAS PERFORMED WHICH REVEALED NO INDICATION OF A PIECE OF THE CATHETER BEING LEFT INSIDE THE PATIENT. PER THE HOSPITAL, NO PATIENT HARM WAS FOUND AND NO FURTHER ACTIONS WERE TAKEN.
Additional Manufacturer Narrative
INVESTIGATION INCLUDED REVIEW OF THE RETURNED DEVICE AND A SAMPLE FROM ANOTHER LOT OF LIKE PRODUCT IN AN ATTEMPT TO REPRODUCE THE FAILURE MODE. THE INVESTIGATION CONCLUDED THAT TENSION APPLIED TO THE DEVICE DURING REMOVAL OF THE CATHETER AGAINST THE SUTURE CAUSED STRETCHING OF THE DISTAL 4CM AND CUTTING OFF OF THE LAST 0.5MM OF THE CATHETER. THE INVESTIGATION INDICATED NO EVIDENCE OF MANUFACTURING RELATED ISSUES THAT COULD HAVE CONTRIBUTED TO THE EVENT.