inforMED
MalfunctionBTM

SUNMED HOLDINGS LLC

Received Jun 1, 2022 · Event occurred Apr 10, 2022

Report 2028807-2022-00014 · MDR key 14569757

Device

Generic name

1 Litre Hyperinflation Bag System W/ Manometer

Model number

HS4011

Catalog number

HS4011

Lot number

210700765

Product problems

  • Manufacturing, Packaging or Shipping Problem

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE, HOWEVER THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. IF NOT CORRECTED IMMEDIATELY PATIENT WOULD BE IMPACTED AND SERIOUS HARM COULD OCCUR. BASED ON THE REPORTED INFORMATION THE CRITERIA FOR REPORTING AN ADVERSE EVENT HAVE BEEN MET.

Description of Event or Problem

THE COMMITTEE ON TECHNOLOGY FOR THE ANESTHESIA PATIENT SAFETY FOUNDATION RECEIVED A REPORT REGARDING OUR HYPERINFLATION SYSTEM DESCRIBING THE DESIGN AS COUNTER-INTUITIVE AND THAT IT DOES NOT CONFORM WITH ISO STANDARD RECOMMENDATIONS. A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE, HOWEVER THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. THE PROBLEM WAS REMEDIED BY QUICK OBSERVATION THAT TO CLOSE THE VALVE, THE VALVE MUST BE TURNED ANTI-CLOCKWISE.