MalfunctionFHO

PNEUMOPERITONEUM NEEDLE

Received Jan 15, 1998

Report 2939738-1997-00042 · MDR key 145198

Device

Generic name

Pneumoperitoneum Needle

Model number

OMS-N120

Catalog number

0MS-N120

Lot number

UNK

Product problems

  • Retraction Problem
  • Device Issue

Patient

UNKNOWN

  • Other (for use when an appropriate patient code cannot be identified)

Narrative

Additional Manufacturer Narrative

COMPLAINT CONFIRMED. THE NEEDLE STYLET DOES NOT RETRACT BACK WHEN THE NEEDLE IS PUSHED FORWARD. CAUSE OF FAILURE: THE CAUSE OF FAILURE WAS DUE TO MISALIGNMENT BETWEEN THE HOUSING ASSEMBLY AND THE STYLET. THE MISALIGNMENT INCREASES THE FRICTION BETWEEN THE NEEDLE AND THE STYLET CAUSING THE STYLET NOT TO RETRACT. IS CORRECTIVE OR PREVENTIVE ACTION REQUIRED? YES. HAS ACTION ALREADY BEEN TAKEN OR ISSUED FOR A SIMILAR EVENT? YES.

Description of Event or Problem

DURING A LAPAROSCOPIC PROCEDURE, THE REPORTER INDICATED THAT THE SPRING ON THE NEEDLE ALLEGEDLY IS NOT WORKING.

Remedial action

  • Recall