SPINDLE NUT
Received May 26, 2022
Report 2939274-2022-01944 · MDR key 14517435
Device
Generic name
Apparatus, Traction, Non-powered
Manufacturer
Wrights Lane Synthes Usa Products LlcModel number
394.42Catalog number
394.42
Product problems
- Device-Device Incompatibility
- Device-Device Incompatibility
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT NUT-KNURLED F/LARGE-DISTRACT WAS STUCK ON THE RETURNED MATING DEVICE. A DIMENSIONAL INSPECTION WAS UNABLE TO BE PERFORMED DUE TO THE COMPLAINT CONDITION. A FUNCTIONAL TEST WAS PERFORMED AND THE COMPLAINT DEVICE WAS UNABLE TO BE REMOVED FROM THE MATING DEVICE. THE COMPLAINT CONDITION WAS ABLE TO BE REPLICATED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE NUT-KNURLED F/LARGE-DISTRACT WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THEY HAVE A SYNTHES THREADED ROD AND A WASHER SLEEVE THAT IS ATTACHED FIRMLY. THE OPERATING ROOM DEPT WAS NOT ABLE TO REMOVE THIS WASHER FROM THE ROD. THIS REPORT INVOLVES ONE (1) SPINDLE NUT. THIS IS REPORT 1 OF 2 FOR (B)(4).