RAPTOR GRASPING DEVICE
Received May 26, 2022 · Event occurred Feb 9, 2022
Report 1528319-2022-00027 · MDR key 14514849
Device
Generic name
Grasping Device
Manufacturer
United States Endoscopy Group, Inc.Model number
00711177Catalog number
00711177
Lot number
2109708
Product problems
- Difficult to Open or Close
- Difficult to Open or Close
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
STERIS HAS MADE MULTIPLE ATTEMPTS TO CONTACT THE USER FACILITY REGARDING THE EVENT ASSOCIATED WITH THE MEDWATCH HOWEVER, THE USER FACILITY HAS NOT RESPONDED. THE INSTRUCTIONS FOR USE PACKET (IFU 731747) LISTS WARNINGS AND PRECAUTIONS WHEN USING THE RAPTOR GRASPING DEVICE, ACTUATE THE DEVICE BY MOVING THE SLIDER ON THE HANDLE BACK AND FORTH TO CONFIRM THAT THE GRASPING JAWS OPEN AND CLOSE SMOOTHLY. IF THE UNIT DOES NOT FUNCTION PROPERLY, OR THERE IS EVIDENCE OF DAMAGE (E.G. BENDS, KINKS, MISSHAPEN JAWS, MISALIGNED JAWS, EXPOSED WIRES) DO NOT USE THIS PRODUCT AND CONTACT YOUR LOCAL PRODUCT SPECIALIST. DO NOT USE EXCESSIVE FORCE ON THE HANDLE AND DO NOT COIL THE CATHETER OUTSIDE OF THE ENDOSCOPE. EXCESSIVE FORCE OR COILING MAY DAMAGE THE DEVICE OR DAMAGE THE ENDOSCOPE AND MAY RESULT IN ACCIDENTAL INJURY TO THE PATIENT OR CLINICIAN. THE FOLLOWING CONDITIONS MAY CAUSE THE DEVICE TO FUNCTION IMPROPERLY: ADVANCING THE HANDLE TO THE OPEN POSITION WITH TOO MUCH SPEED OR FORCE. ATTEMPTING TO PASS OR OPEN THE DEVICE IN AN EXTREMELY ARTICULATED ENDOSCOPE. ATTEMPTING TO ACTUATE THE DEVICE IN AN EXTREMELY COILED POSITION. ACTUATING THE DEVICE WHEN THE HANDLE IS AT AN ACUTE ANGLE IN RELATION TO THE SHEATH. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND CONFIRMED THE DEVICE LOT WAS MANUFACTURED TO SPECIFICATION. A COMPLAINT REVIEW CONFIRMS THE LOT SUBJECT OF THE REPORTED EVENT TO BE ISOLATED. A FOLLOW-UP MDR WILL BE SUBMITTED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.
Description of Event or Problem
THE USER FACILITY REPORTED VIA MEDWATCH # 5201030000-2022-8002 THAT DURING A PATIENT PROCEDURE THEIR RAPTOR GRASPING DEVICE WOULD NOT OPEN. USER FACILITY PERSONNEL UTILIZED AN ADDITIONAL RAPTOR GRASPING DEVICE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY OR PROCEDURE DELAYS.