inforMED
MalfunctionJDP

INTERLOCKING BOLT FOR 4.5MM VA-LCP CONDYLAR INSERTN HANDLE

Received May 23, 2022

Report 8030965-2022-03405 · MDR key 14470077

Device

Generic name

Orthopaedic Implant Aiming Arm

Manufacturer

Synthes Gmbh

Catalog number

03.231.005

Lot number

L463664

Product problems

  • Device-Device Incompatibility
  • Device-Device Incompatibility

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. REPORTER IS A J&J SALES REPRESENTATIVE. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE INTERLOCKING BOLT FOR 4.5 VA-LCP CURVED CONDYLAR PLATE INSERTION HANDLE (03.231.005) GOT A DEFECT AT THE DISTAL THREAD. ACCORDING TO THE REPORT, THIS DEFECT CAUSED A CROSS THREAD IN THE VA-LCP CURVED CONDYLAR PLATE (02.124.415) DURING INSERTION OF THE INTERLOCKING BOLT HANDLE IN THE PLATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DELAY OF FIVE MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS COMPLAINT INVOLVES TWO (2) DEVICES.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D2: ADDITIONAL PROCODES: HRS, HWC. D10. H6: PART: 03.231.005-US. LOT: L463664. MANUFACTURING SITE: HAGENDORF. SUPPLIER: N/A. RELEASE TO WAREHOUSE DATE: 02 AUGUST 2017. EXPIRATION DATE: N/A. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. THE PHOTO WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE DEVICE FROM THE RECEIVED IMAGE(S). VISUAL ANALYSIS OF THE PHOTO REVEALED THAT THE BOLT F/AIMING ARM INSTR F/VA-LCP CONDYLA WAS FOUND IN THE EVIDENCE PROVIDED IT LOOKS THE TIP OF THE DEVICE WORN. NO OTHER DEFECT WAS OBSERVED. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED FOR THE BOLT F/AIMING ARM INSTR F/VA-LCP CONDYLA. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. D1.

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: H3, H6: PART: 03.231.005-US, LOT: L463664, MANUFACTURING SITE: HAGENDORF, SUPPLIER: N/A, RELEASE TO WAREHOUSE DATE: 02 AUG 2017, EXPIRATION DATE: N/A. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT BOLT F/AIMING ARM INSTR F/VA-LCP CONDYLA SHOWS THE DISTAL THREADS ARE STRIPPED/WORN, PREVENTING THE DEVICE FROM FUNCTIONING AN INTENDED. A DIMENSIONAL INSPECTION FOR THE BOLT F/AIMING ARM INSTR F/VA-LCP CONDYLA WAS UNABLE TO BE PERFORMED DUE TO POST MANUFACTURING DAMAGE. A FUNCTIONAL TEST WAS PERFORMED WITH THE MATING DEVICE (PLATE) AND CANNOT BE CONNECTED CORRECTLY DUE TO OBSERVED CONDITION DURING THE VISUAL INSPECTION. THE COMPLAINT CONDITION WAS ABLE TO BE REPLICATED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE BOLT F/AIMING ARM INSTR F/VA-LCP CONDYLA WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.