inforMED
MalfunctionFTC

SPECTRA UV THERAPY BOOTH

Received Jun 28, 1994 · Event occurred Apr 11, 1994

Report MW1002622 · MDR key 14371

Device

Generic name

Uv Therapy Booth

Manufacturer

Daavlin Co.

Model number

305/350

Product problems

  • Nonstandard Device
  • Output, high

Patient

47 YR

Narrative

Description of Event or Problem

UNIT FAILED TO TERMINATE AT PRESET TIME AND MILIJOULES ACCORDING TO PT TREATMENT PLAN. PT REPORTED MACHINE WAS GETTING HOT AND THOUGHT HE'D BEEN IN THERE TOO LONG. METER READ 8:11 MINUTES BUT STILL 190 MILIJOULES LEFT. PHYSICIAN EXAMINED PT AND NO APPARENT SYMPTOMS AT THAT TIME. DURING INVESTIGATION IT WAS FOUND THAT THE UNIT HAD AN FDA RECALL IN 1991, FOR THE SAME PROBLEM. THAT WAS 2 YRS BEFORE RPTR PURCHASED UNIT. IT APPEARS THE MFR SOLD RPTR A UNIT THAT WASN'T IN FDA COMPLIANCE.