MalfunctionPXE

MEDEX LOGICAL PRESSURE MONITORING SYSTEM

Received May 11, 2022 · Event occurred Jan 15, 2022

Report 3012307300-2022-08578 · MDR key 14365170

Device

Generic name

Pressure Monitoring (air/gas) Kit

Catalog number

MX960P1CZ

Lot number

4227294

Product problems

  • High Readings
  • Low Readings
  • High Readings
  • Low Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE MONITOR READING DIFFER BETWEEN N1BP AND 1BP IS OUT OF RANGE. IBP VALUE IS DIFFERENT FORM THE N1BP VALUE.

Additional Manufacturer Narrative

A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED WHICH INDICATED ALL INSPECTIONS WERE COMPLETED AND NO ISSUES WERE NOTED DURING MANUFACTURE. A PRODUCT SAMPLE WAS RECEIVED FOR EVALUATION. VISUAL AND FUNCTIONAL TESTING WERE PERFORMED. VISUAL INSPECTION FOUND NO PHYSICAL MARKINGS OR DAMAGE NOTED ON THE RETURNED UNIT. FUNCTIONAL TESTING FOUND THE DEVICE MEETS SPECIFICATIONS. DEVICE PASSED ALL FUNCTIONAL TESTING. NO PROBLEM FOUND. ROOT CAUSE CANNOT BE DETERMINED AS THE SAMPLE WAS SUCCESSFULLY TESTED AND NO DISCREPANCIES WERE DETECTED.