16 FR. FLEXIBLE VIDEO CYSTOSCOPE
Received May 11, 2022
Report 1221826-2022-00075 · MDR key 14355470
Device
Generic name
Flexible Video Cystoscope
Manufacturer
Karl Storz EndovisionModel number
11272VNUKCatalog number
11272VNUK
Product problems
- Contamination /Decontamination Problem
- Contamination /Decontamination Problem
Patient
Unknown
- Unspecified Infection
- Unspecified Infection
Narrative
Additional Manufacturer Narrative
DEVICE WAS RETURNED FOR EVALUATION. PER THE MANUFACTURER'S EVALUATION FINDINGS: PER THE TESTING RESULTS, NO LINK COULD BE MADE TO THE PATIENT INFECTIONS AND ANY PERSISTENT MICROBIAL GROWTH (I.E. BIOFILM) COULD NOT BE IDENTIFIED. IT WAS FOUND THE SHAFT IS CRUSHED 48 MM FROM SHAFT ADAPTER AND THE VERTEBRAE SYSTEM IS CRUSHED 30 MM FROM DISTAL TIP. UNABLE TO CONFIRM CUSTOMER'S COMPLAINT.
Description of Event or Problem
THE FLEXIBLE CYSTOSCOPE WAS USED IN 4 CASES AND THE CLINIC REPORTS 2 PATIENTS CONTRACTED AN INFECTION. THIS REPORT IS FOR THE SECOND OF THE TWO CASES. REPORT #1221826-2020-00059 WAS PREVIOUSLY SUBMITTED FOR ONE OF THE TWO CASES.