VITAL SYNC REMOTE MONITORING SYSTEM
Received May 9, 2022 · Event occurred Apr 22, 2022
Report 1717344-2022-00557 · MDR key 14338126
Device
Generic name
Transmitters And Receivers, Physiological Sign
Manufacturer
Covidien Mfg Dc BoulderModel number
10116080Catalog number
10116080
Product problems
- Computer Software Problem
- Computer Software Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ACCORDING TO THE REPORTER, THE REMOTE MONITORING SYSTEM HAD WINDOWS SCRIPT ERROR AND WAS UNABLE TO SET ALARMS IN THE SERVER. THE DEVICES WERE ALARMING IN THE ROOM BUT ARE NOT SEEN IN THE SERVER. THE SCRIPT ERROR WAS NOTED TO FREEZE OR CLOSE OUT THE SOFTWARE. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative
EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE IS STILL IN USE AND WAS EVALUATED BASED ON TECH NICAL SERVICE OR SUPPORT. VISUAL INSPECTION NOTED THE ALARM SYSTEM WAS NOT SETUP CORRECTLY FOR THE VITAL SYNC SYSTEM. THE ISSUE WAS RESOLVED BY TURNING DOWN THE VOLUME OF THE MACHINE. IT WAS REPORTED THAT THE VITAL SYNC SYSTEM DID NOT PRODUCE ALARM MESSAGES. THE REPORTED ISSUE WAS CONFIRMED. THE MOST LIKELY CAUSE WAS TRACED TO SOFTWARE CODING. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.