MalfunctionPCI

BD MAX¿ ENTERIC BACTERIAL PANEL

Received May 5, 2022 · Event occurred Mar 23, 2022

Report 3007420875-2022-00019 · MDR key 14289452

Device

Generic name

Gastrointestinal Bacterial Panel Multiplex Nucleic Acid-based Assay System

Model number

442963

Catalog number

442963

Lot number

1074231

Product problems

  • False Negative Result
  • False Negative Result

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem

IT HAS BEEN REPORTED THAT THE BD MAX¿ ENTERIC BACTERIAL PANEL HAS BEEN FOUND PRODUCING FALSE NEGATIVE RESULTS. CULTURES WERE USED FOR CONFIRMATORY TESTING. DELAYED PATIENT TREATMENT WAS REPORTED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS INVESTIGATING A SUSPECTED PROBLEM WITH ENTERIC BACTERIAL PANEL ON THE BD MAX AND ITS ABILITY TO DETECT SALMONELLA. IN THE LAST 4-5 WEEKS CUSTOMER HAS HAD 3 BD MAX SALMONELLA NEGATIVE RESULTS THAT ARE CULTURE POSITIVE. TWO OF THESE ARE S. TYPHI AND ONE WAS DETECTED OVER A WEEK AFTER THE INITIAL NEGATIVE RESULT.

Description of Event or Problem

CUSTOMER IS INVESTIGATING A SUSPECTED PROBLEM WITH ENTERIC BACTERIAL PANEL ON THE BD MAX AND IT HAS BEEN REPORTED THAT THE BD MAX¿ ENTERIC BACTERIAL PANEL HAS BEEN FOUND PRODUCING FALSE NEGATIVE RESULTS. CULTURES WERE USED FOR CONFIRMATORY TESTING. DELAYED PATIENT TREATMENT WAS REPORTED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS INVESTIGATING A SUSPECTED PROBLEM WITH ENTERIC BACTERIAL PANEL ON THE BD MAX AND ITS ABILITY TO DETECT SALMONELLA. IN THE LAST 4-5 WEEKS CUSTOMER HAS HAD 3 BD MAX SALMONELLA NEGATIVE RESULTS THAT ARE CULTURE POSITIVE. TWO OF THESE ARE S. TYPHI AND ONE WAS DETECTED OVER A WEEK AFTER THE INITIAL NEGATIVE RESULT.

Additional Manufacturer Narrative

H6: INVESTIGATION SUMMARY: THE COMPLAINT INVESTIGATION FOR DISCREPANT RESULTS WHEN USING THE BD MAX¿ ENTERIC BACTERIAL PANEL ASSAY (REF. 442963) LOT 1074231 WAS PERFORMED BY THE REVIEW OF THE MANUFACTURING RECORDS, RETAIN MATERIAL TESTING AND BY THE COMPLAINT¿S HISTORY REVIEW. REVIEW OF THE MANUFACTURING RECORDS OF BD MAX¿ ENTERIC BACTERIAL PANEL ASSAY INDICATED THAT THE LOT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS AND MET PERFORMANCE REQUIREMENTS. THE RETAIN MATERIAL OF BD MAX¿ ENTERIC BACTERIAL PANEL FROM LOT 1074231 WAS TESTED IN POSITIVE AND ALL THE REACTIONS GAVE POSITIVE RESULTS, AS EXPECTED. CUSTOMER REPORTED THREE NEGATIVE SAMPLES WITH BD MAX¿ ENTERIC BACTERIAL PANEL WHICH GAVE A POSITIVE RESULT FOR SALMONELLA IN CULTURE. CUSTOMER STATED THAT TWO OF THEM WERE S. TYPHI AND ONE OF THEM WAS DETECTED ONE WEEK AFTER THE INITIAL NEGATIVE RESULT. DESPITE SEVERAL REQUESTS MADE TO THE CUSTOMER, NO DATA OR ADDITIONAL INFORMATION WAS PROVIDED FOR THE INVESTIGATION. SEQUENCE ANALYSIS REVEALED THAT SALMONELLA TYPHI IS DETECTED WITH THE BD MAX¿ ENTERIC BACTERIAL PANEL ASSAY. MULTIPLE CAUSES COULD EXPLAIN THE CUSTOMER REPORTED DISCREPANT RESULTS, BUT WITHOUT DATA NO ANALYSIS WAS POSSIBLE. ALTHOUGH THE EXACT CAUSE FOR THE CUSTOMER ISSUE COULD NOT BE IDENTIFIED, BASED ON THE INVESTIGATION THE REAGENTS ARE NOT SUSPECTED. THE ROOT CAUSE WAS NOT IDENTIFIED. THERE IS NO INDICATION OF A REAGENT ISSUE BASED ON THE ANALYSIS OF THE COMPLAINTS RECEIVED FOR DISCREPANT RESULTS ON BD MAX¿ ENTERIC BACTERIAL PANEL LOT 1074231. BD CANNOT CONFIRM THE COMPLAINT BASED ON THE INVESTIGATION THAT WAS PERFORMED. BD DID NOT INITIATE A CORRECTIVE AND PREVENTIVE ACTION (CAPA). BD APOLOGIZES FOR THE INCONVENIENCE THAT THIS MAY HAVE CAUSED. BD QUALITY WILL CONTINUE TO MONITOR FOR TRENDS.