inforMED
MalfunctionIZZ

DEFINIUM 8000

Received Apr 22, 2009 · Event occurred Mar 20, 2009

Report 2126677-2009-00029 · MDR key 1424813

Device

Generic name

Kpr/izf/mqb

Model number

5131070

Catalog number

NA

Lot number

NA

Product problems

  • Unintended System Motion

Patient

NA

  • No Patient Involvement

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE TABLE LOCKS DISENGAGED WITHOUT FOOT PEDAL ACTIVATION, CAUSING AN UNEXPECTED FREE FLOAT MOTION OF THE TABLETOP IN THE LONGITUDINAL AND LATERAL DIRECTIONS. THE SITE INDICATED THAT THE FLOAT OCCURRED AS THE TABLE WAS LOWERED TO ITS LOWEST SETTING. THERE WAS NO PT INVOLVEMENT AND NO INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING OR UNLOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.

Additional Manufacturer Narrative

GE FIELD ENGINEER (FE) FOUND THAT THE FOOT PEDAL MECHANISM WAS NOT ALIGNED PROPERLY. THE FE READJUSTED THE PEDAL AND VERIFIED THE TABLE LOCKS WERE PERFORMING AS EXPECTED.