inforMED
MalfunctionHTT

ULTRAPOWER BUR, ROUND, 1.0MM

Received Apr 23, 2009 · Event occurred Apr 2, 2009

Report 1017294-2009-00062 · MDR key 1423512

Device

Generic name

Instrument, Surg, Orthopedic, Pneumatic

Manufacturer

Conmed Linvatec

Model number

NA

Catalog number

00702122200

Lot number

BBC41339

Product problems

  • Loose or Intermittent Connection

Patient

NA

  • No Patient Involvement

Narrative

Description of Event or Problem

FOREIGN COUNTRY'S DISTRIBUTOR REPORTED THAT DURING RECEIVING AND INSPECTING OF THIS DEVICE, THEY FOUND THE BUR CAME LOOSE FROM THE CLAMSHELL INSIDE THE STERILE PACKAGING. THERE WAS NO REPORT OF COMPROMISE TO THE STERILITY OF THE PRODUCT. THERE WAS NO PT INVOLVEMENT, INJURY OR SURGICAL DELAY RESULTING FROM THIS EVENT.

Additional Manufacturer Narrative

INVESTIGATION FINDINGS: CONMED LINVATEC RECEIVED THIS BUR AND PACKAGING FOR EVAL. IN 2009, A VISUAL EXAMINATION OF THE PACKAGE CONFIRMED THE REPORT OF THE BUR LOOSE FROM THE CLAMSHELL INSIDE THE STERILE PACKAGING. FURTHER INVESTIGATION FOUND SCRATCHES ON THE POUCH AND CONFIRMED A COMPROMISE TO THE STERILITY OF THE PACKAGING. CONMED LINVATEC INITIATED AND COMPLETED A CORRECTIVE AND PREVENTIVE ACTION TO ADDRESS THIS PROBLEM. A RETAINER COMPONENT WAS ADDED TO THE PACKAGING TO PREVENT THE BUR FROM COMING LOOSE INSIDE THE CLAMSHELL.