inforMED
MalfunctionIZZ

COMPAX 40E

Received Apr 1, 2009 · Event occurred Mar 6, 2009

Report 2126677-2009-00024 · MDR key 1418666

Device

Model number

2133494

Catalog number

NA

Lot number

NA

Product problems

  • Device Slipped
  • Unintended Movement

Patient

NA

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE TABLE LOCKS DID NOT ACTUATE WHEN THE FOOT PEDAL WAS RELEASED, CAUSING THE TABLETOP TO UNEXPECTEDLY MOVE WITHOUT RESISTANCE IN BOTH LONGITUDINAL AND LATERAL DIRECTIONS (FREE FLOAT). THERE WAS NO INJURY REPORTED. THIS SITUATION COULD CONTRIBUTE TO AN INJURY IF A PT OR OPERATOR WERE UNAWARE OF THIS CONDITION WHILE LOADING OR UNLOADING A PT. THE ENSUING INSTABILITY COULD LEAD TO A FALL.

Additional Manufacturer Narrative

THE GE FIELD ENGINEER (FE) FOUND THAT THE UNEXPECTED FLOAT WAS A RESULT OF A POWER LOSS TO THE TABLE LOCKS DUE TO A BLOWN FUSE. THE FE REPLACED THE FUSE AND VERIFIED THAT THE TABLE LOCKS WERE WORKING AS EXPECTED.