inforMED
MalfunctionFII

PHOENIX

Received Jul 1, 2009 · Event occurred Jun 1, 2009

Report 1414156 · MDR key 1414156

Device

Generic name

Hemodialysis Machine

Model number

*

Catalog number

*

Lot number

*

Product problems

  • Leak/Splash
  • Loss of Power

Patient

16 YR

  • No Information

Narrative

Description of Event or Problem

MACHINE SUDDENLY STOPPED DURING A TREATMENT. MACHINE HAD A GENERAL SAFE STATE ALARM THAT ATTRIBUTED TO FAILURE OF THE PI/PO FAILURE OF THE T1 TEST. THE CAUSE OF THE FAILURE WAS TRACED TO A LEAK PRESENT IN THE PUF PERISTALTIC PUMP SEGMENT FAILURE. CHANGED THE PUMP SEGMENT AND T1 TEST PASSED FOUR CONSECUTIVE TIMES.