MalfunctionFBO

INNER CERAMIC INSULATION TUBE

Received Apr 12, 2022 · Event occurred Mar 9, 2022

Report 9610617-2022-00031 · MDR key 14084349

Device

Generic name

Resectoscope Sheath, Inner Fixed Sheath

Model number

27040XB

Catalog number

27040XB

Lot number

NV01

Product problems

  • Break
  • Material Separation
  • Break
  • Material Separation

Patient

Male

  • Foreign Body In Patient
  • Foreign Body In Patient

Narrative

Additional Manufacturer Narrative

THE CERAMIC TIP FRACTURED/BROKE FROM THE SHEATH AT END OF PROCEDURE WHEN PULLING OUT INSTRUMENT. A MISSING PIECE IS STILL IN PATIENT. THE SHEATH IS OVER 8 YEARS OLD.

Description of Event or Problem

DURING A TURP, SURGEON COMPLETED THE SURGERY, AND REALIZED WHEN PULLING OUT THE INSTRUMENTS THAT A BROKEN PIECE OF CERAMIC CAME OUT. A MISSING PIECE IS STILL IN THE PATIENT.

Additional Manufacturer Narrative

AN INVESTIGATION REPORT FROM KST/MFR. WAS RECEIVED WHICH STATES: THE CONCLUSION AND ROOT CAUSE OF THE NON-RETURN PRODUCT: "THE MOST LIKELY CAUSE IS THE AGE ( 8 YEARS) OF THE ITEM AND THE HIGH" APPLICATION CYCLES. THIS PROBABLY CAUSED MICRO CRACKS ON THE CERAMIC BEAK, WHICH LED TO THE BREAKAGE OF THE BEAK DURING EXTRACTION.