Received Apr 1, 2022 · Event occurred Mar 4, 2022

Report 1823260-2022-00928 · MDR key 13975880

Device

Generic name

Assay, Glycosylated Hemoglobin

Manufacturer

Roche Diagnostics

Model number

HBA1C TQ G3

Catalog number

05336163190

Lot number

ASKU

Product problems

  • High Test Results
  • High Test Results

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE INITIAL REPORTER QUESTIONED HIGH RESULTS FOR 2 SAMPLES FROM 1 PATIENT TESTED FOR A1C-3 ON A COBAS 8000 C 502 MODULE. ON APPROXIMATELY (B)(6) 2022 THE RESULT FROM THE C502 MODULE WAS 31.8%. THIS RESULT WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS CONSIDERED TO BE NOT BELIEVABLE. THIS SAMPLE WAS REPEATED BY AN UNSPECIFIED CHROMATOGRAPHY METHOD AND THE RESULT WAS 17.8%. THIS RESULT WAS BELIEVED TO BE CORRECT. ON (B)(6) 2022 A 2ND SAMPLE WAS OBTAINED AND THE RESULT FROM THE C502 MODULE WAS 28.7%. SOMETIME DURING THE WEEK OF (B)(6) 2022 ANOTHER SAMPLE WAS OBTAINED FROM THE PATIENT AND THE RESULT FROM THE C502 MODULE WAS 17.1% WHICH CORRESPONDS TO THE RESULT FROM THE CHROMATOGRAPHY METHOD. ON (B)(6) 2022 THE CUSTOMER REPEATED ANOTHER SAMPLE FROM THE PATIENT ON THE C502 MODULE AND THE RESULT WAS 16.8% WHICH CORRESPONDS TO THE RESULT FROM THE CHROMATOGRAPHY METHOD. THE C502 MODULE SERIAL NUMBER WAS (B)(4).

Additional Manufacturer Narrative

NO FURTHER INFORMATION OR DATA WAS PROVIDED BY THE CUSTOMER. SAMPLE MATERIAL WAS REQUESTED FOR INVESTIGATION BUT WAS NO LONGER AVAILABLE TO SEND. FROM THE INFORMATION PROVIDED, THE CAUSE OF THE EVENT COULD NOT BE DETERMINED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.