DEVICE, CYSTOMETRIC, HYDRAULIC
Received Mar 31, 2022 · Event occurred Mar 1, 2022
Report 3017448360-2022-00010 · MDR key 13968999
Device
Generic name
Device, Cystometric, Hydraulic
Manufacturer
Laborie Medical TechnologiesModel number
CAT, TDOC, 1X SENS., COU. TIP, 7 FR 7FSCCatalog number
CAT878
Lot number
214195
Product problems
- Device Dislodged or Dislocated
Patient
Unknown
- Abrasion
- Hemorrhage/Blood Loss/Bleeding
- Abrasion
- Hemorrhage/Blood Loss/Bleeding
Narrative
Description of Event or Problem
THE EVENT INVOLVED A TDOC CATHETER, MODEL NUMBER CAT878, USED FOR A URODYNAMIC STUDY. DURING THE STUDY THE PATIENT NOTIFIED THE PROVIDER THAT THE CATHETER WAS UNCOMFORTABLE, AND THE CATHETER WAS REMOVED. THE PATIENT EXPERIENCED BLEEDING WHICH WAS CONTROLLED WITH GAUZE AND PRESSURE APPLICATION. UPON EXAMINATION OF THE CATHETER USED, THE SENSOR WIRE WITHIN THE CATHETER HAD DISLODGED FROM THE END OF THE CATHETER LIKELY CAUSING AN ABRASION OF THE URETERAL MUCOSA. THE PATIENT HAS URINARY RETENTION REQUIRING INTERMITTENT CATHERIZATION AND AN INDWELLING CATHETER HAD TO BE PLACED AT THE TIME TO AVOID FURTHER IRRITATION OF LIKELY ABRASION CAUSED BY THE CATHETER. THIS INCIDENT DID NOT RESULT IN A SERIOUS INJURY OF THE PATIENT. THE EVENT IS BEING REPORTED AS IF THIS INCIDENT WAS TO OCCUR AGAIN MAY REQUIRE MEDICAL INTERVENTION TO PREVENT SERIOUS INJURY OF THE PATIENT.