M220 F12
Received Mar 29, 2022 · Event occurred Mar 5, 2022
Report 3003974370-2022-00003 · MDR key 13936028
Device
Generic name
Surgical Microscope
Manufacturer
Leica Microsystems (schweiz) AgModel number
M220 F12Catalog number
10448367
Product problems
- Device Fell
- Device Fell
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS IS A COMBINED INITIAL/FINAL REPORT. LEICA MICROSYSTEMS PERFORMED A ROOT CAUSE INVESTIGATION WHICH IDENTIFIED THE CAUSE OF THE OBSERVED PROBLEM AS INSUFFICIENT SECURING OF THE SCREW CONNECTING THE SWING ARM TO THE OPTICS CARRIER. IT WAS DETERMINED THAT THIS SCREW WAS NOT ADEQUATELY SECURED WITH THE APPROPRIATE AMOUNT OF HIGH STRENGTH THREAD ADHESIVE TO ENSURE A RELIABLE CONNECTION OF THE SWING ARM TO THE OPTICS CARRIER. THIS INSUFFICIENT SECURING OF THE SWING ARM WAS NOT IDENTIFIED DURING PRODUCTION AND FINAL INSPECTION OF THE M220 F12 SWING ARM. THESE SWING ARMS WERE THEN BUILT INTO THE M220 F12 SURGICAL MICROSCOPES MANUFACTURED BEFORE MARCH 9TH, 2017. AS A RESULT, THE SCREW THAT CONNECTS THE SWING ARM TO THE OPTICS CARRIER HAS LOOSENED OVER TIME AND CONSEQUENTLY CAUSED THE OPTICS CARRIER TO FALL DOWN. BASED ON THE RESULTS OF THE ROOT CAUSE INVESTIGATION, THE PROBABILITY OF THE END USER DETECTING THIS DEFECT IS VERY LOW. CAPA-(B)(4) AND THE ASSOCIATED FSCA (FDA RECALL #: Z-0412-2022; FDA RECALL ID: 89009) WERE INITIATED TO ADDRESS THIS ISSUE. THE DEFECTIVE SWING ARM WAS REPLACED BY A NEW ONE.
Description of Event or Problem
LEICA MICROSYSTEMS (SCHWEIZ) AG RECEIVED A COMPLAINT FROM CHINA STATING THAT M220 OPTICS CARRIER FELL DOWN WHEN PREPARING FOR THE NEXT SURGERY. THERE WAS NO PATIENT OR USER INJURY.
Remedial action
- Recall