InjuryFSE
SCHUREMED BEACH CHAIR
Received Mar 25, 2022 · Event occurred Feb 11, 2022
Report MW5108525 · MDR key 13928678
Device
Generic name
Table, Operating-room, Manual
Manufacturer
Schuerch CorporationProduct problems
- Defective Component
Patient
Unknown
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
Narrative
Description of Event or Problem
OPERATING ROOM PATIENT WAS ON TRUMPF 7000U TABLE WITH SCHUREMED SHOULDER CHAIR. WHEN STARTING TO POSITION PATIENT, CHAIR BECAME UNLOCKED AND UNLATCHED FROM THE TABLE WHICH CAUSED THE PATIENT TO BEND OVER BACKWARDS. INVESTIGATION REVEALED DEFECTIVE LOCKING MECHANISM ON SHOULDER CHAIR, WHICH THEN CAUSED IT TO BECOME UNLATCHED FROM THE TABLE. FDA SAFETY REPORT ID # (B)(4).