inforMED
MalfunctionJDP

INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL

Received Mar 26, 2022 · Event occurred Feb 25, 2022

Report 2939274-2022-01070 · MDR key 13915470

Device

Generic name

Condylar Plate Fixation Implant

Model number

03.231.005

Catalog number

03.231.005

Lot number

2L14077

Product problems

  • Material Twisted/Bent
  • Material Twisted/Bent

Patient

55 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT ON (B)(6) 2022, WHILE PERFORMING A DISTAL FEMUR FRACTURE, THE INTERLOCKING BOLT STRIPPED OUT AND WOULD NOT ALLOW THE HANDLE TO BE ATTACHED TO THE PLATE. A SEPARATE INTERLOCKING BOLT WAS USED, AND THE SURGERY PROCEEDED. THERE WAS A SURGICAL DELAY OF 5 MINUTES. NO PATIENT CONSEQUENCES. THIS REPORT IS FOR ONE (1) INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10 ADDITIONAL NARRATIVE: D4, D9 H3, H6: A PRODUCT INVESTIGATION WAS CONDUCTED. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED SAMPLE REVEALED THAT THERE WAS NO DAMAGE OR DEFECTS WITH THE INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL. A DIMENSIONAL INSPECTION FOR THE INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL WAS UNABLE TO BE PERFORMED AS IT IS IRRELEVANT TO THE REPORTED COMPLAINT CONDITION. A FUNCTIONAL TEST COULD NOT BE PERFORMED AS THE MATING DEVICE WAS NOT RETURNED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE INTERLCKNG BOLT 4.5 VA COND INSERTN HNDL WAS FOUND TO HAVE NO DAMAGE OR DEFECTS. NO DEFINITIVE ROOT CAUSE COULD BE DETERMINED. THERE WAS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE CONTRIBUTED TO THE COMPLAINT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. H3, H4, H6: A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED: PART # 03.231.005-US LOT # 2L14077 MANUFACTURING SITE: WERK HAGENDORF RELEASE TO WAREHOUSE DATE: 05 DEC2018 A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.