inforMED
MalfunctionLEJ

ASTRAL 150 - APAC1

Received Mar 18, 2022 · Event occurred Feb 16, 2022

Report 3007573469-2022-00179 · MDR key 13812268

Device

Manufacturer

Resmed Ltd

Model number

27083

Catalog number

27083

Product problems

  • Defective Device

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED TO RESMED AND AN EVALUATION CONFIRMED THE COMPLAINT. THE POWER SUPPLY UNIT WAS REPLACED TO ADDRESS THIS ISSUE. THE DEVICE WAS SERVICED AND FULLY TESTED BEFORE IT WAS RETURNED TO THE CUSTOMER. RESMED REFERENCE #: (B)(4).

Description of Event or Problem

IT WAS REPORTED TO RESMED THAT AN ASTRAL POWER SUPPLY UNIT (PSU) WAS INOPERABLE. THERE WAS NO PATIENT HARM OR SERIOUS INJURY REPORTED AS A RESULT OF THIS INCIDENT.