inforMED
MalfunctionMZR

LEUKOSTRIP 13X102MM CTN 50

Received Mar 16, 2022 · Event occurred Feb 24, 2022

Report 8043484-2022-00102 · MDR key 13777974

Device

Generic name

Topical Approximation System

Model number

66002880

Catalog number

66002880

Lot number

2053A

Product problems

  • Contamination
  • Contamination

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A BLACK STAIN IN THE STRIP OF LEUKOSTRIP 13X102MM CTN 50. AS THIS WAS NOTICED DURING SET UP, THE TREATMENT WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT IMPACTED.

Additional Manufacturer Narrative

NTERNAL COMPLAINT REFERENCE: (B)(4).

Additional Manufacturer Narrative

H3, H6: THE PRODUCT HAS BEEN RETURNED AND EVALUATED, THE EVENT IS CONFIRMED. VISUAL INSPECTION OF THE DRESSING HAS CONFIRMED THE REPORTED CONTAMINATION, THIS HAS BEEN CONFIRMED TO BE ADDITIONAL ADHESIVE, THIS IS NOT ANY OTHER FORM OF CONTAMINATION. A DOCUMENTATION REVIEW HAS REVEALED PREVIOUSLY CASES WITH A SIMILAR NATURE, THERE ARE NO OPEN CORRECTIVE ACTIONS. THE MANUFACTURING RECORDS CONFIRM THE DEVICE MET THE FINAL PRODUCT SPECIFICATIONS AND CONTAIN NO CONTRIBUTORY FACTORS TO THE REPORTED EVENT, THE ASSOCIATED RISK FILES MITIGATE THIS EVENT TYPE. A REVIEW OF THE MANUFACTURING PROCESS HAS REVEALED THE PROBABLE CAUSE. A PROBABLE CAUSE INCLUDES THE DEFECT WAS NOT PICKED OUT FROM RAW MATERIAL PRIOR TO PRODUCTION, RETRAINING HAS BEEN CONDUCTED, TO REDUCE THE PROBABILITY OF FUTURE OCCURRENCES. NO CORRECTIVE ACTIONS ARE DEEMED NECESSARY, AT THIS STAGE, SMITH AND NEPHEW WILL CONTINUE TO MONITOR FOR ADVERSE TRENDS RELATING TO THE REPORTED EVENT.