inforMED
No answer providedFLL

DRAGER RADIANT WARMER BED TEMPERATURE PROBE

Received Mar 8, 2022 · Event occurred Dec 24, 2021

Report 13691863 · MDR key 13691863

Device

Generic name

Thermometer, Electronic, Clinical

Product problems

  • Low Readings

Patient

Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

NOTICED INFANT WAS WARM IN AFTERNOON CARE. RN TRIED TO WEAN THE BED TEMPERATURE DOWN. INFANT WAS WARMER AT NEXT CARE TIME. RN ATTEMPTED TO WEAN BED TEMPERATURE A SECOND TIME. RN CHECKS TEMP AN HOUR LATER, TEMPERATURE IS 101.4 AND TEMP PROBE IS ON CORRECTLY BUT READING COLD AND BED IS AT HIGHEST HEAT INTENSITY. PRODUCT SAVED FOR PRODUCTS NURSE. MANUFACTURER AWARE OF ISSUE. NEW PROBE USED, ISSUE DID NOT REPEAT.