No answer providedFLL
DRAGER RADIANT WARMER BED TEMPERATURE PROBE
Received Mar 8, 2022 · Event occurred Dec 24, 2021
Report 13691863 · MDR key 13691863
Device
Generic name
Thermometer, Electronic, Clinical
Manufacturer
Draegerwerk Ag & Co. KgaaProduct problems
- Low Readings
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
NOTICED INFANT WAS WARM IN AFTERNOON CARE. RN TRIED TO WEAN THE BED TEMPERATURE DOWN. INFANT WAS WARMER AT NEXT CARE TIME. RN ATTEMPTED TO WEAN BED TEMPERATURE A SECOND TIME. RN CHECKS TEMP AN HOUR LATER, TEMPERATURE IS 101.4 AND TEMP PROBE IS ON CORRECTLY BUT READING COLD AND BED IS AT HIGHEST HEAT INTENSITY. PRODUCT SAVED FOR PRODUCTS NURSE. MANUFACTURER AWARE OF ISSUE. NEW PROBE USED, ISSUE DID NOT REPEAT.