HTP1500
Received Mar 4, 2022 · Event occurred Dec 17, 2020
Report 13666185 · MDR key 13666185
Device
Generic name
Pack, Hot Or Cold, Water Circulating
Manufacturer
Adroit Medical Systems, Inc.Model number
HTP1500Catalog number
HTP1500
Product problems
- Device Contamination with Chemical or Other Material
Patient
8395 DA · Female
- Unspecified Infection
- Unspecified Infection
Narrative
Description of Event or Problem
THE PATIENT UNDERWENT ECMO PLACEMENT FOR CIRCULATORY SUPPORT AFTER BEING ADMITTED IN CRITICALLY ILL CONDITION IN (B)(6) 2020. THE ECMO DEVICE WAS PAIRED WITH A DISPOSABLE WATER HEATER TO WARM THE ECMO CIRCUIT. THE PATIENT REMAINED ON ECMO FOR HER ENTIRE HOSPITAL COURSE FROM (B)(6) 2020 THROUGH (B)(6) 2022. THE PATIENT EVENTUALLY UNDERWENT LUNG AND KIDNEY TRANSPLANTS IN (B)(6) 2021. THE PATIENT DEVELOPED INFECTION IN HER LUNGS WITH CUPRIAVIDUS. THE WATER HEATER WAS CULTURED IN (B)(6) 2021 WHICH REVEALED RALSTONIA, PSEUDOMONAS, CUPRIAVIDUS AND SPHINGOMONAS. DESPITE AGGRESSIVE TREATMENT, THE PATIENT EXPIRED IN (B)(6) 2022.