inforMED
MalfunctionDWZ

ENDOMYOCARDIAL BIOPSY FORCEP

Received Mar 1, 2022 · Event occurred Jan 25, 2022

Report 13633932 · MDR key 13633932

Device

Generic name

Device, Biopsy, Endomyocardial

Model number

190030

Catalog number

190030

Lot number

5180239

Product problems

  • Break
  • Difficult to Remove

Patient

16060 DA · Male

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

THE BIOPSY FORCEP IS A LONG CATHETER THAT HAS A HANDLE ON THE END, THE HANDLE IS CONNECTED TO THE FORCEPS AT THE OTHER END OF THE CATHETER BY A WIRE INSIDE THE CATHETER, THIS ALLOWS THE FORCEPS TO BE OPENED AND CLOSED INSIDE THE HEART FROM OUTSIDE THE BODY. PART OF THIS MECHANISM BROKE, MAKING IT DIFFICULT TO CONTROL THE OPENING AND CLOSING OF THE FORCEPS. THIS, IN TURN, MADE REMOVAL OF THE FORCEPS FROM THE BODY VIA THE SHEATH DIFFICULT. A HEMOSTAT WAS USED TO CONTROL THE DEVICE IN A WAY THAT FACILITATED THE REMOVAL OF THE DEVICE SUCCESSFULLY.

Description of Event or Problem

THE BIOPSY FORCEP IS A LONG CATHETER THAT HAS A HANDLE ON THE END, THE HANDLE IS CONNECTED TO THE FORCEPS AT THE OTHER END OF THE CATHETER BY A WIRE INSIDE THE CATHETER, THIS ALLOWS THE FORCEPS TO BE OPENED AND CLOSED INSIDE THE HEART FROM OUTSIDE THE BODY. PART OF THIS MECHANISM BROKE, MAKING IT DIFFICULT TO CONTROL THE OPENING AND CLOSING OF THE FORCEPS. THIS, IN TURN, MADE REMOVAL OF THE FORCEPS FROM THE BODY VIA THE SHEATH DIFFICULT. A HEMOSTAT WAS USED TO CONTROL THE DEVICE IN A WAY THAT FACILITATED THE REMOVAL OF THE DEVICE SUCCESSFULLY.