INFANT HEEL WARMER,SQUEEZE, 4X4IN
Received Feb 24, 2022 · Event occurred Jan 31, 2022
Report 1423537-2022-00721 · MDR key 13595484
Device
Generic name
Infant Heel Warmer (chemical Heat Pack)
Manufacturer
Mo016 Mo-moberlyModel number
11460-010TCatalog number
11460-010T
Lot number
V1P119
Product problems
- Burst Container or Vessel
- Burst Container or Vessel
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
CUSTOMER REPORTED THAT WHILE A BLOOD DRAW WAS BEING PERFORMED IN AN OUTPATIENT LAB, A TEAM MEMBER ATTEMPTED TO USE AN INFANT HEEL WARMER. THE TEAM MEMBER POINTED THE PACK AWAY FROM PATIENT AND MEMBER AND ACTIVATED THE WARMER BY SQUEEZING. THE PACK SQUIRTED OUT FLUID, BUT IT DID NOT LAND ON PATIENT OR ON TEAM MEMBER. THERE WAS NO INJURY REPORTED.
Additional Manufacturer Narrative
BASED ON SUPPLIER INVESTIGATION, THE DEVICE HISTORY RECORD REVIEW WAS COMPLETED ON THE REPORTED LOT NUMBER, V1P119. THE LOT WAS FOUND TO HAVE BEEN MANUFACTURED AND RELEASED TO PREDETERMINED SPECIFICATIONS WITH NO ANOMALIES FOUND. ONE SAMPLE WAS RECEIVED ON 2/15/2022 THAT WAS ALREADY ACTIVATED. THE ROOT CAUSE WAS DETERMINED TO BE TOP SEAL RUNOFF DUE TO INADEQUATE PROCEDURE. THE SITE HAS UPDATED THE PROCEDURE FOR THE OPERATORS TO ENSURE THEY ARE VERIFYING & DOCUMENTING THAT THE SYSTEM IS WORKING PROPERLY ON EACH MACHINE DURING MANUFACTURING. ADDITIONALLY, THE MANUFACTURING PLANT WILL CONTINUE TO MONITOR THIS ISSUE THROUGH ROUTINE COMPLAINT TREND ANALYSIS.