HIGH SPEED ELEC G1 HANDPIECE AIR COOLED
Received Feb 24, 2022
Report 1045834-2022-00210 · MDR key 13591410
Device
Generic name
Motor, Drill, Electric
Manufacturer
Depuy Synthes Products LlcModel number
EG1ACatalog number
EG1A
Product problems
- Excessive Heating
- Excessive Heating
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, AND IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY. UDI: (B)(4).
Description of Event or Problem
IT WAS REPORTED THAT THE MOTOR DEVICE WAS HEATING UP. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED AS APPROPRIATE. H10 ADDITIONAL NARRATIVE: DEVICE EVALUATION: THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING REPAIR, IT WAS DETERMINED THAT THE REPORTED CONDITION OF MOTOR DEVICE HEATING UP WAS NOT CONFIRMED. THEREFORE, AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. HOWEVER, DURING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD MISSING COMPONENTS AND FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION. IT WAS FURTHER DETERMINED THAT THE DEVICE COULD NOT ENGAGE ATTACHMENT. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR ATTACHMENT ASSESSMENT. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE TRACED TO MAINTENANCE, WHICH IS IMPROPER MAINTENANCE.